SCATTER RADIATION

Testing Summary

 

STARTable

Background

Adept Medical STARTable Shield is embedded with 0.5mm Lead (Pb), offering protection from scatter radiation for operators. To verify this, testing was carried out to measure the product’s effectiveness in a controlled environment, by simulating scatter radiation exposure to operators and measuring x-ray dose around a gridded pattern.

The testing was conducted at Auckland District Health Board Interventional Cardiology Laboratory’s on a Siemens Artis Imaging Table. The measurements were recorded by Brian Lunt, Medical Physicist. A radiographic human torso phantom was used to simulate a patient on the table and provides the primary source of radiation to the clinician and staff via scatter. 

Results

At gantry tilt 0°

  • The maximum percentage scatter radiation reduction throughout the grid is 90%.

  • At 110 cm (waist height – 4D correlating to usual clinician position during procedure), the scatter radiation reduction with use of Adept Medical STARTable is 77%.

  • At 150 cm (neck height – 6D correlating to usual clinician position during procedure), the scatter radiation reduction with use of Adept Medical STARTable is 65%.

  • At the cranial end of the STARTable, the scatter radiation reduction is negligible.

  • At the caudal end of the STARTable, the scatter radiation reduction is 79%.

At LAO 34°/CRAN 20° 

  • The maximum percentage scatter radiation reduction throughout the grid is 89%.
     

  • At 110 cm (waist height – 4D correlating to usual clinician position during procedure), the scatter radiation reduction with use of Adept Medical STARTable is 87%.
     

  • At 150 cm (neck height – 6D correlating to usual clinician position during procedure), the scatter radiation reduction with use of Adept Medical STARTable is 64%.
     

  • At the cranial end of the STARTable, the scatter radiation reduction is negligible.
     

  • At the caudal end of the STARTable, the scatter radiation reduction is 89%.

 

X-Ray Shield

Background

Adept Medical X-Ray Shield (X-Ray Shield) is embedded with 0.5mm Lead (Pb), offering protection from scatter radiation for operators. To verify this, testing was carried out to measure the product’s effectiveness in a controlled environment to simulate scatter radiation exposure around a gridded pattern.

The testing was conducted at Auckland District Health Board Interventional Cardiology Laboratory’s on a Siemens Artis Imaging Table. The measurements were recorded by Brian Lunt, Medical Physicist. A radiographic human torso phantom was used to simulate a patient on the table and provides the primary source of radiation to the clinician and staff via scatter. 

Results

At gantry tilt 0°

  • The maximum scatter radiation reduction throughout the grid is 93%.

  • At 110 cm (waist height – 4D, correlating to usual clinician position during procedure), the scatter radiation reduction with use of Adept Medical X-Ray Shield is 93%.
     

  • At 150 cm (neck height), the maximum scatter radiation reduction with use of Adept Medical X-Ray Shield is 68%.
     

  • At 150 cm (neck height – 6D) , the scatter radiation reduction with use of Adept Medical X-Ray Shield is 41%.

At RAO 34°/CRAN 20° 

  • The maximum attenuation throughout the grid is 85%.

  • At 110 cm (waist height), the maximum scatter radiation reduction with use of Adept Medical X-Ray Shield is 85%.
     

  • At 110 cm (waist height – 4D, correlating to usual clinician position during procedure), the scatter radiation reduction with use of Adept Medical X-Ray Shield is 78%.
     

  • At 150 cm (neck height), the maximum scatter radiation reduction with use of Adept Medical X-Ray Shield is 61%.
     

  • At 150 cm (neck height – 6D, correlating to usual clinician position during procedure) , the scatter radiation reduction with use of Adept Medical X-Ray Shield is 56%.